Parameter | Specification |
---|---|
HPLC column temperature | 30°C |
HPLC flow rate | 1.5 mL/min |
HPLC analysis time | 10 min |
HPLC injection volume | 5 µL |
HPLC mobile phase conditions | Mobile phase A: aqueous ammonium formate (5mM)/formic acid (0.1%) Mobile phase B: acetonitrile/formic acid (0.1%) Isocratic, 25% mobile phase A |
GC-MS carrier gas | Hydrogen at 1 mL/min |
GC-MS inlet temperature | 200°C |
GC-MS injection volume: | 1 µL at 50:1 split ratio |
GC-MS temperature programme | 60°C and held for 2 min, increasing to 300°C at 8°C/min and held for 10 min |
GC-MS ion source temperature | 230°C |
GC-MS mass analyzer temperature | 150°C |
GC-MS analysis time | 42 min |
GC-MS scan range | 40–600 m/z at 1 Hz |