Skip to main content
Fig. 3 | Journal of Cannabis Research

Fig. 3

From: THC degradation does not impair the accuracy of THC doses aerosolized by the metered-dose SyqeAir inhaler: a 24-month stability trial

Fig. 3

Amount of major phytocannabinoids in MC stored in Syqe cartridges at room temperature over 2 years. The middle circle represents mean value and the upper and lower lines represent the upper and lower 95% confidence interval values. BL, baseline; M, month; CBDA, cannabidiol acid; THC, (-)-∆9-trans- tetrahydrocannabinol; THCA, (-)-∆9-trans- tetrahydrocannabinol acid; CBGA, cannabigerolic acid; CBN, cannabinol; “Total” represents the sum of the acid and the neutral components. Statistically significant differences between times of storage were calculated by one-way ANOVA, utilizing Fisher’s exact test analyses (whenever significant, stated as the p value and when not significant stated as N.S.), followed by a Tukey post hoc multiple comparison test (*, p<0.05; **, p<0.01; ***, p<0.001)

Back to article page